HZ-7080 Tracking Index Tester
HZ-7080 Tracking Index Tester
OVERVIEW

Product summary:
The instrument uses a rectangular platinum electrodes, bipolar force on the sample were 1.0N ± 0.05N. Applied voltage is adjustable between 100 ~ 600V (48 ~ 60Hz), short-circuit current 1.0 ± 0.1A, and the voltage drop should not when more than 10%. When the test circuit, short-circuit current is equal to or greater than the leakage 0.5A, the time is for 2 seconds, the relay to cut off the current, indicating that the test failed. Adjustable time constant drip device, precise control of the amount of the liquid droplet size of 44 ~ 50 drops / cm3 and the drip interval 30 ± 5 seconds.
 
Meet the standards:
ASTM D3638, GB/T 4207-2003
 
Product Application:
The tester can measure the voltage up to 600V when solid electrical insulating materials in the electric field on the surface exposed to relatively resistant to tracking performance when water containing impurities.
 
The main technical requirements:

Electrode 2mm × 5mm sectional area of the two rectangular platinum electrodes. End edge of 30 ° slope.
The surface forces 1.0±0.05N
Testing voltage 100~600V
The maximum test current 3A
Two electrode distance 4.0mm
Drip device Drip interval can be set
Volume of the chamber 0.5M3, L*W*H: 60*95*90cm
Dimension L*W*H: 61*120*105 cm
External material Electrostatic paint and stainless steel mirror. According to customer requirements and business case selected negotiated