HZ-1708B Thermal deformation
HZ-1708B Thermal deformation
OVERVIEW

Equipment description:
HZ-1708B  Thermal deformation, Vicat softening temperature testermainly used for the thermal deformation of the polymer materials (plastics, rubber, nylon, electric insulating materials, etc.) and Vicat softening point temperature determination. Apply to universities, research institutes and Product Quality Supervision and Inspection unit.Heated media inside the 45 ° double helix automatic mixing system, the tank has a special structure, temperature uniformity, and maintain the accuracy of ± 0.5 ° C. And the use of advanced thyristor-controlled temperature, displacement sensor has a fine-tuning, fine tuning the dual adjustment method,with emergency stop switch, the structure of the locked door after the Control Panel, machine maintenance, easy maintenance, the lift system as one of the models to reduce the noise. Therefore, the medium temperature gradient is uniform, constant temperature limit is reached when the test temperature to stop heating automatically when the set temperature, the test frame with automatic lift function, simple operation, with excellent cost performance in similar models.

Meet the standards:
GB / T 1633-2000 (idt ISO 306:1994) Vicat softening temperature (VST) determination
GB / T 1634.1-2004 (idt ISO75-2: 2003) plastic load deformation temperature measurement (Common test methods)
GB / T 1634.2-2004 plastic load deformation temperature measurement (plastic, hard rubber, and long fiber reinforced composites)
GB / T 1634.3-2004 plastic load deformation temperature determination (high-strength thermosetting laminates)
GB / T 8802-2001 (eqv ISO 2507:1995) thermoplastic pipes, fittings Determination of Vicat softening temperature

Technical parameters:
Temperature range: room temperature -300 ℃
Heating rate: 120 ° C / h, 50 ° C / h
The maximum temperature measurement error: +0.5 ° C
The maximum temperature control error: +12 ℃ / min (heat distortion test)+5 ° C (Vicat test) / 6   minutes
Heat deflection: Table 1, Table 2
The Vicat change deflection: 1mm
The maximum deformation measurement error: +0.005 mm
The number of sample holder: 3
A weight to weight (ie, load lever, standard head, the sensor probe, trays, screws): 68g +1 g
Heating medium: methyl silicone oil (200 cSt below the flash point of 300 ° C and above)
Maximum heating power: 3kw
Cooling: Natural cooling or water cooling
Main Power: 220V +10%, 15A, 50HZ;

Weight:

         Name   Specification A B C D E F G H I J K
Quality(g) 1000 932 500 200 100 50 20 10 5 2 1
number 12 3 3 6 3 3 6 3 3 6 3